Original Article

Volume: 13 | Issue: 1 | Published: Dec 10, 2023 | Pages: 20 - 23 | DOI: 10.24911/PJNMed.175-1692263460

Estimation of Net Percent Error in Radioiodine-131 Activity during Administration to Patients


Authors: Javaid Ali , Shoab Shah , Abdul Samad , Imadullah Tariq , Ghufran Biradr , Muhammad Sohail , Irum Naz Memon , Attia Gul , Israr Ahmad


Article Info

Authors

Javaid Ali

Larkana Institute of Nuclear Medicine & Radiotherapy (LINAR) hospital, Larkana, Pakistan.

Shoab Shah

Swat Institute of Nuclear Medicine, Oncology & Radiotherapy (SINOR) hospital, Swat, Pakistan.

Abdul Samad

Larkana Institute of Nuclear Medicine & Radiotherapy (LINAR) hospital, Larkana, Pakistan.

Imadullah Tariq

Larkana Institute of Nuclear Medicine & Radiotherapy (LINAR) hospital, Larkana, Pakistan.

Ghufran Biradr

Swat Institute of Nuclear Medicine, Oncology & Radiotherapy (SINOR) hospital, Swat, Pakistan.

Muhammad Sohail

International Collaborative Laboratory of 2D Materials for Optoelectronics Science and Technology of Ministry of Education, Institute of Microscale Optoelectronics, College of Electronics and Information Engineering, Shenzhen University, Shenzhen, 518060, China.

Irum Naz Memon

Larkana Institute of Nuclear Medicine & Radiotherapy (LINAR) hospital, Larkana, Pakistan.

Attia Gul

Institute of Nuclear Medicine, Oncology & Radiotherapy (INOR) hospital, Abbottabad, Pakistan.

Israr Ahmad

Larkana Institute of Nuclear Medicine & Radiotherapy (LINAR) hospital, Larkana, Pakistan.

Publication History

Received: August 17, 2023

Accepted: October 05, 2023

Published: December 10, 2023


Abstract


Background: Radiation workers have reported that some errors during measurement of radioiodine-131 activities have been observed, which may result poor quality images, greater radioactive waste and financial burden in the hospital. The purpose of this retrospective study was to estimate the net percent errors in radioiodine-131 activities during administration to thyroid patients and give some fruitful suggestions to reduce such errors.

Methods: We included180 random patients i.e. thirty for each advised activity of 3 mCi, 15 mCi, 20 mCi, 25 mCi, 30 mCi and 150 mCi in this study from January 2019 to June 2022. All the recommended activities were measured with CRC-25R dose calibrator. The percent difference in the measured & prescribed activities and percent difference in the measured & residual activities were estimated. Finally, the net percent errors in administered activities were measured for all patients.

Results: The net percent error in recommended activities of 3 mCi, 15 mCi, 20 mCi, 25 mCi, 30 mCi & 150 mCi and administered activities were [6.73 (-32.7, 20.95) ± 11.15] %, [3.91 (-3.85, 8.77) ± 2.94] %, [3.63 (-2.39, 7.79) ± 2.51] %, [3.55 (-0.65, 10.67) ±2.6] %, [2.93 (-1.98, 6.78) ± 1.89] % & [0.88 (-1.04, 2.68) ±0.87] % respectively.

Conclusion: This study concludes that the patients receive approximately 6.73%, 3.91%, 3.63 %, 3.55 %, 2.93 % and 0.88 % less activities than the aforementioned activities respectively. The net percent error in radioiodine-131 activities is not alarming, but needs to be adjusted in order to ensure high quality images and optimum dose delivery.


Keywords: Radioiodine-131, residual activity, thyroid disorder, thyroid carcinoma